Measurement functions |
Number of channels: 2, for distortion measurement (electronic auto-balancing, balance adjustment range within ±10000 με) |
Input connectors |
Via adapter cable, TAJIMI PRC03-12A10-7M10.5, Max. rated voltage to earth:
33 Vrms or 70 V DC (with input isolated from the unit, the maximum voltage that can be
applied between input channel and chassis and between input channels without damage) |
Suitable
transducer |
Strain gauge converter, bridge impedance: 120 Ω to 1 kΩ (bridge voltage 2 V),
350 Ω to 1 kΩ (bridge voltage 5 V, 10 V), bridge voltage 2, 5, 10 ±0.05 V |
Measurement
range |
20 με - 1000 με/division, 6 ranges, full scale: 20 divisions, lowpass
filter: 5/10/100/1 kHz |
Anti-aliasing filter |
Integrated filter for suppressing aliasing distortion caused by FFT
processing (automatic cutoff frequency setting/OFF) |
Measurement resolution |
1/1600 of measurement range (using 16-bit A/D conversion; installed in 8860/8861) |
Highest sampling rate |
200 kS/s (2-channel simultaneous sampling) |
Accuracy After auto-balancing |
DC amplitude: ±(0.4% of full scale +2 με), zero position: ±0.1% of full scale
(at 5 Hz filter ON) |
Frequency characteristics |
DC to 20 kHz +1/-3 dB |
Max. allowable input |
10 V DC (the maximum voltage that can be applied across input pins without damage) |