| DC Current | 20.00 A/600.0 A, Basic accuracy: ±1.3% rdg. ±0.08 A (20 A range) |
| AC Current | 20.00 A/600.0 A (10 Hz to 1 kHz, True RMS), Basic accuracy: ±1.3% rdg. ±0.08 A (at 20 A) |
| Crest factor | 20.00 A range: 7.5 600.0 A range: 3 or less |
| DC+AC Current | 20.00 A/600.0 A (10 Hz to 1 kHz, True RMS), Basic accuracy:±1.3% rdg. ±0.13 A (at 20 A) |
| DC Voltage | 600.0 mV to 1500 V, 5 ranges, Basic accuracy: ±0.5% rdg. ±3 dgt. |
| AC Voltage | 6.000 V to 1000 V, 4 ranges (15 Hz to 1 kHz, True RMS), Basic accuracy: ±0.9% rdg. ±3 dgt. |
| DC+AC Voltage | 6.000 V to 1000 V, 4 ranges, Basic accuracy: ±1.0% rdg. ±7 dgt. |
| Resistance | 600.0 Ω to 600.0 kΩ, 4 ranges, Basic accuracy: ±0.7% rdg. ±5 dgt. |
| Electrostatic capacity | 1.000 μF to 1000 μF, 4 ranges, Basic accuracy: ±1.9% rdg. ±5 dgt. |
| Frequency | 9.999 Hz to 999.9 Hz, Basic accuracy: ±0.1% rdg. ±1 dgt. |
| Temperature (K) | -40.0 to 400.0 ˚C, add temperature probe accuracy to basic accuracy of ±0.5% rdg. ±3.0 ˚C |
| Voltage detection | Hi: 40 V to 600 V AC, Lo: 80 V to 600 V AC, 50/60 Hz |
| Other functions | Automatic AC/DC detection, Plus/Minus judgement function of DC A, DC V, Max/Min/Average/PEAK MAX/ PEAK MIN value display, Filter function, Display value hold, Auto hold, Back light, Auto-power save, Buzzer sounds, Zero-adjustment |
| Dustproof, waterproof | Grip: IP54 (when measuring an insulated conductor only), Jaw (the current sensor portion of the instrument)/ barrier: IP50 *Risk of electric shock from the conductor being measured increases when wet. |
| Power supply | LR03 Alkaline battery ×2, Continuous use 45 hr (Backlight OFF) |
| Core jaw diameter | 69 mm (2.72 in) W× 14 mm (0.55 in) D, φ33 mm (1.30 in) |
| Dimensions and mass | 65 mm (2.56 in) W × 215 mm (8.46 in) H × 35 mm (1.38 in) D mm, 340 g (12.0 oz) |
| Accessories | TEST LEAD L9207-10, CARRYING CASE C0203, LR03 Alkaline battery ×2, Instruction Manual×1 |