| Sensor | CT7046 | CT7045 | CT7044 |
| Rated measurement current | 6000A AC |
| Internal Measurement range | 600A AC/ 6000A AC (Range is controlled by main device) |
| Max. allowable input | 10000 A continuous (at 6000 A range, 45 to 66 Hz, requires derating ) |
| Bandwidth | 10 Hz to 50 kHz (±3dB) |
| Amplitude and phase accuracy | ±1.5 % rdg. ±0.25 % f.s. (f.s. is internal range, 45 to 66 Hz) ±1 deg |
| Output rate | 1 mV/A (600 A), 0.1 mV/A (6000 A) |
| Max. rated voltage to earth | AC 600 V (CAT IV), AC 1000 V (CAT III) |
| Loop diameter | φ 254 mm (10.00 in) or less | φ 180 mm (7.09 in) or less | φ 100 mm (3.94 in) or less |
| Dustproof waterproof | IP54* (When sensor is connected to a compatible instrument.) * Do not use when wet. |
| Output connectors | HIOKI PL 14 |
| Dimensions | Flexible loop cable diameter: φ7.4 mm (0.29 in), Cable length: Between flexible loop and battery box: 2.3 m (7.55 ft), Output cable: 20 cm (0.66 ft), Battery box25 mm (0.98 in) W × 72 mm (2.83 in)H × 20 mm (0.79 in)D |
| Mass | 186 g (6.6 oz) | 174 g (6.1 oz) | 160 g (5.6 oz) |
| Accessories | N/A |